For the measurement of pressures fluctuating
above and below atmospheric, and critical test work.
ALL PLASTIC
LARGE BORE COLUMNS
TOUGH YET LIGHT
FIVE SIZES TO COVER ALL REQUIREMENTS
AVAILABLE AS SEVERAL UNITS ON ONE PANEL
CONSTRUCTION Crystal clear plastic tubes, O-sealed into solid clear blocks.
White scale, engraved black, adjustable for zero, mounted between
the columns on the same plane as the bores to prevent parallax errors.
Stout backing board.
TYPES C/UT AND D2/UT Solid Perspex with
no backing boards.
Readings taken from one column only.
RANGES Five sizes standardised with metre scales. Available with imperial
and S.l. metric scales to special order.
Liquid Required
C/UT
250-0.250 mm
25cc
D2/UT
500-0-500 mm
25cc
UT 5
500-0-500 mm
sub division 2 mm
25 cc
UT 10
1000-0-1000 mm
sub division 2 mm
50 cc
UT 15
1500-0-1500 mm
sub division 2 mm
75cc
UT 20
2000-0-2000 mm
sub division 2 mm
100 cc
INDICATING LIQUIDS Liquid S.G.1.0, coloured blue-for
direct reading in mm H20
Mercury-triple distilled - for direct reading in mm Hg.
Micro-porous discs are supplied to insert under the connectors
to prevent any loss of mercury, while allowing air or gas to pass.
Inert liquid S.G.1.9. approx, coloured BLUE - for reading in mm
H2O, multiply by exact S.G. given
on bottle of liquid. e.g. true pressure mm H2O
= reading x 1.972. Gives greater range than water, but much greater
sensitivity than mercury. Useable temperature range -20°C to
40°C.
ACCESSORIES VALVES -- Isolating, equalising or
venting valves are mounted on separate backing board TUBING - - PVC, Nylon CONNECTORS --- Compression type, in place
of 'push-on' supplied as standard RATINGS --- Pressure 50 p.s.i.g. 3 bar.
Temperature 80°C
APPLICATIONS For measurements of pressures, vacuum,
differential pressures, levels and flows.
website
design by:
Poddymeter Limited Unit 2, Park Works, Borough Road, Kingston
upon Thames, Surrey KT2 6BD, United Kingdom Tel: 020 8546 9311 Fax: 020 8547 2325
www.poddymeter.co.uk E-MAIL: