6200MF Hydronic Manometer
Digital Water Flow Test Set for measurement of Flow, Differential Pressure and Static Pressure
Unit Price (Excl. VAT): £1060.00
-
Differential and line pressure in KPa
-
Flow in L/Sec with KVs input
-
Rugged transducer arrangement
-
Sealed unit – water resistant
-
Damped steady readings
-
Overpressure warning
-
Low Battery warning
-
Hanging strap provided
-
Quick Release Tubing with Valves & Binder Adaptor Pair Included
-
Foam lined case for instrument, tubing & adaptor storage
-
Tap adaptor included
-
Calibration certificate
-
Simple operation
-
Back light
Suitable for Flow and Differential Pressure measurement across balancing
valves, orifice plates,Pitot tubes etc. in waterflow systems.
Flow rate can be displayed by inputting individual KVs values.
In ‘Flow’ mode, both flow(in L/Sec) and DP (in KPa) are displayed.
In ‘Pressure mode’ line pressure (in KPa) and DP (in KPa) are both displayed.
All units are leak tested to 10 Bar and calibrated before despatch. The unit
comes complete with a calibration certificate, pair of binder adaptors, nylon
tubing valve/quick release adaptor sets, and a foam lined carry case for
storage.
The instrument is also supplied with a detachable strap for hanging in
awkward areas or around neck.
Simple to use, just connect, flush, turn on, zero, and read. Providing the line
pressure does not exceed 10 Bar, no damage can be incurred by the
transducer arrangement if one side only is accidentally pressurised. If the
static pressure is exceeded, a warning message is displayed. The 6200MF has
quick release connections with robust valves fitted to the tube for easy
storage. The casing is sealed for resistance against water/moisture ingress.
Recommended for hot water systems: A tap adaptor accessory for cold water
priming is available.
Spares tubing & valves for our digital waterflow manometers and the tap adaptor for cold water priming can be found on our commissioning accessories page, along with other commissioning spares such as Binder & Mechseal adaptors.
Specification
(Range:
Line pressure : 0 – 1000 KPa
Resolution 1 Kpa
Differential pressure : 0 – 100.00 KPa
Resolution 0.01 Kpa
Flow : 4 digits in L/Sec
(floating decimal point)
KVs input range : 0.5 - 2500
Max Line Pressure : 10 BAR G
Accuracy:
0 – 20 KPa : +/- 0.1 KPa
Over 20 KPa +/- 0.2 KPa or 0.5% whichever is greater
Batteries : 4 x AA (Supplied)
Battery Life : 100 hours approx
Transducer wetted parts : 316 stainless steel
Instrument Size & weight : 27.5 x 11 x 4 cm ; 750g
Cased Size & weight : 40 x 30 x 9 cm ; 2.4 KG
Calibration Period: 12 Months
Operating Temperature: 0 - 50 deg. C (at transducer)